Volatile liquids like acetone and methanol can be extremely difficult to pipette using air displacement.
Air displacement pipette.
Air displacement pipette used for standard pipetting applications is highly accurate.
Too low too high volume is pipetted.
However conditions such as temperature atmospheric pressure as well as the specific gravity and viscosity of the solution may have an effect on the performance of air displacement pipettes.
Zeus is equipped with anti droplet control adc.
These liquids tend to vaporize in the tip which raises the air pressure resulting in liquid dripping from the tip.
A vacuum is generated by the vertical travel of a metal or ceramic piston within an airtight sleeve.
These pipettes operate by piston driven air displacement.
Due to its high accuracy this laboratory tool is commonly used in standard pipetting applications.
Acids bases and solvents are the worst enemies of any air displacement pipette.
As with any gas the air cushion varies according to the characteristics of liquids as well as in part by lab or.
Adjust pipette to liquid becomes fixed volume pipette for adjusted volume.
In an air displacement pipette an air cushion separates the liquid in the plastic tip from the piston inside the pipette.
When enabled the pressure is monitored in the tip.
Unlike an air displacement pipette which dispenses liquid using an air cushion in the pipette tip the piston in a positive displacement pipette makes direct contact with the sample allowing the aspiration force to remain constant.
Getting down to the basics an air displacement pipette is a common laboratory tool used to handle a measured volume of liquid between 1 µl to 1000 µl 1 ml.
When using positive displacement pipettes there is no elastic air space to expand or contract so the aspiration force remains constant unaffected by the physical properties of the sample.
Laboratory pipettes can function either via air displacement or by positive displacement principles.
Positive displacement pipettes are a type of pipette that operates via piston driven displacement.
Both pipette types have a piston that moves in a cylinder or capillary.
In positive displacement pipetting the piston is in direct contact with the liquid.
No air cushion therefore no problem.